Gamma camera

Gamma and fluorescence camera for ionizing radiation imaging (prompt gamma, X-ray fluorescence) with spectroscopic voxel detector. Two layer configuration of detectors (Si and CdTe sensors) for increased efficiency of the camera. Two modes of operation - camera obscura and Comton camera (particle track reconstruction).

Technical parameters:
sensor area: 2x 1.98 cm2
resolution: 460 DPI (pixel pitch 55um)
operation modes:   collimated - camera obscura
           particle track reconstruction – Compton camera
sensors:  1000 um CdTe
      300 um Si
energy range: 3-1300 keV (photons and charged particles)
readout speed: 30 fps for both layers
PC connection via USB
Software support: Pixelman software package (Win/Linux/Mac)

gamma_camera_3 gamma_camera_4 gamma_camera_1 gamma_camera_2